Download Planar.prc |
PalmOS PDA Applications:PlanarA popular basis for brachytherapy QA are the Paterson-Parker implant tables. These tables were originally designed for radium implants according to the Manchester system (Meredith WJ, ed. Radium dosage. The Manchester System. Edinburgh: Livingston 1967). The data in the original tables gave the number of mg hours required to deliver an exposure of 1000 roentgens (mg-h/1000R). These data were based on a value of 0.84 for the exposure rate constant (Γ) for radium. It is common for the original tables to be modified to yield mg-h/10Gy by multiplying the original values by the factor C = 0.84 / (0.825 * 0.957) = 1.064 which corrects for the alteration of Γ for radium and converts exposure to dose to tissue. The tables can be adapted for HDR use by converting the original values to a more convenient form for Ir-192 (Γ = 0.469) HDR by multiplying the original values by C and then again by H = (0.001 * 3600) * (0.825 / 0.469) = 6.333 to yield mCi-sec/cGy. The HDR dwell time in seconds is calculated by multiplying the tabulated mCi-sec/cGy values by the prescribed dose (in cGy) and dividing by the current activity of the source (in mCi). Click here to download the .zip archive: Planar 1.0.0 (20K), last updated 9/8/03. Download: Planar.prc.zip (16K)
|